Evaluation of the illuminance on the monitor screen present in the hemodynamics room

Authors

DOI:

https://doi.org/10.33448/rsd-v11i8.31172

Keywords:

Hemodynamics; Diagnostic Imaging; Radiological Protection; Lighting; Ocular adaptation.

Abstract

Objective: To evaluate, using a photometer, the influence of luminance from ambient illuminance reflected on the monitor screen on the visualization of low-contrast structures during hemodynamic procedures. Method: To simulate a low-contrast situation, some PMMA plates were carefully arranged and exposed to radiation, both in normal and low-dose fluoroscopy modes, available in the device. A photometer was used to measure the luminance on the monitor screen and measurements were performed for 12 different lighting scenarios in the hemodynamics room. Results: It was possible to observe from the luminance measurements performed in each scenario, that the positioning of the luminaires can influence the visualization of low-contrast structures on the monitor screen, and that the fact that the luminaires are on or partially off in the hemodynamics room, can be an important factor to ensure the visual comfort of the physician responsible for the procedure.

References

ABNT (2013). Associação Brasileira de Normas Técnicas. Norma Brasileira ISO/CIE 8995-1. Iluminação de ambientes de trabalho Parte 1: Interior.

Adamsson, M., Laike, T., Morita, T. (2018). Comparison of static and ambulatory measurements of illuminance and spectral composition that can be used for assessing light exposure in real working environments. LEUKOS, 15(2-3), 181-194.

Baker, P. A., Raos, A. S., Thompson, J. M., & Jacobs, R. J. (2013). Visual acuity during direct laryngoscopy at different illuminance levels. Anesthesia & Analgesia, 116(2), 343-350.

Bouffard, M. A. (2019). The Pupil. Continuum (Minneapolis, Minn.), 25(5), 1194-1214.

Brasil (2022). Ministério da Saúde. Diretoria Colegiada da Agência Nacional de Vigilância Sanitária. Resolução RDC Nº 611, de 9 de março de 2022. Estabelece os requisitos sanitários para a organização e o funcionamento de serviços de radiologia diagnóstica ou intervencionista.

Brasil (2021). Ministério da Saúde. Diretoria Colegiada da Agência Nacional de Vigilância Sanitária. Instrução Normativa IN Nº 91, de 27 de maio de 2021. Estabelece os requisitos sanitários para a garantia da qualidade e da segurança de sistemas de fluoroscopia e de radiologia intervencionista.

Bushberg, J. T., Seibert, J. A., Leidholdt Jr., E. M., & Boone, J. M. (2001). The essential physics of medical imaging. 2a ed. Lippincott Williams & Wilkins.

Chen, A. H., Norazman, F. N. N., & Buari, N. H. (2012). Comparison of visual acuity estimates using three different letter charts under two ambient room illuminations. Indian Journal of Ophthalmology, 60(2), 101.

Durán, A., Hian, S. K., Miller, D. L., Le Heron, J., Padovani, R., & Vano, E. (2013). Recommendations for occupational radiation protection in interventional cardiology. Catheterization and Cardiovascular Interventions, 82(1), 29-42.

Estrela, C. (2018). Metodologia científica: ciência, ensino, pesquisa. 3a ed. Artes Médicas.

Francisco, M. F. F., de Camargo, R. B. F., Hirayama, T., Rocha, I. G., & Pires, S. R. (2020). Avaliação do contraste-detalhe em sistemas digitais de equipamentos de hemodinâmica. Revista Brasileira de Física Médica, 14, 556.

Gubolino, L. A., Lopes, M. A. C. Q., Pedra, C. A. C., Caramori, P. R. A., Mangione, J. Á., Silva, S. S., ... & Barbosa, M. R. (2013). Diretrizes da sociedade brasileira de cardiologia sobre qualidade profissional e institucional, centro de treinamento e certificação profissional em hemodinâmica e cardiologia intervencionista. Arquivos Brasileiros de Cardiologia, 101(6), 1-58.

Guyton, A. C., & Hall, J. E. (2006). Tratado de Fisiologia Médica. 11a ed. Guanabara Koogan.

IAEA (2018). International Atomic Energy Agency. Radiation protection and safety in medical uses of ionizing radiation.

Krupinski, E. A., Berbaum, K. S., & Caldwell, R. (2009). Impact of visual fatigue on observer performance. In Medical imaging 2009: image perception, observer performance, and technology assessment (Vol. 7263, p. 72631O). International Society for Optics and Photonics.

Losso, E., Costa, T. D., Carbente, R. P., Barros, F. S., & Nohama, P. (2013). Iluminação em ambientes médico-hospitalares. VI Simpósio em Engenharia Biomédica da Universidade Federal de Uberlândia.

Mesquita, B. R., Sousa, D. N., & dos Santos, L. A. P. (2018). Dispositivo alternativo para uso em controle de qualidade de monitores de imagem empregados no radiodiagnóstico médico. Scientia Plena, 14(10).

NCRP (2010). National Council on Radiation Protection and Measurements. Radiation dose management for fluoroscopically guided interventional medical procedures.

Romani, T. K., & Nogueira, M. C. D. J. A. (2013). Avaliar a iluminação artificial nos centros cirúrgicos em Cuiabá: estudo de caso. Revista Eletrônica em Gestão, Educação e Tecnologia Ambiental, 9(9), 2037-2048.

Rose, A. (1948). The sensitivity performance of the human eye on an absolute scale. JOSA, 38(2), 196-208.

R Core Team (2022). R: A language and environment for statistical computing. R Foundation for Statistical Computing.

Simonson, E., & Brozek, J. (1948). The effect of spectral quality of light on visual performance and fatigue. JOSA, 38(10), 830-840.

Stone, P. T., Clarke, A. M., & Slater, A. I. (1980). The effect of task contrast on visual performance and visual fatigue at a constant illuminance. Lighting Research & Technology, 12(3), 144-159.

Taylor, B. N. The International System of Units (SI) [Internet] (1998). National Institute of Standards and Technology.

Tobarra-López, A., Seguí-Crespo, M., García Bernabeu, J. R., & Boj, P. G. (1995). Calibrado del programa CSF-Test utilizado como sistema de medida de la función de sensibilidad al contraste.

Vetter, C., Pattison, P. M., Houser, K., Herf, M., Phillips, A. J. K., Wright, K. P., Skene, D. J., Brainard, G. C., Boivin, D. B., & Glickman, G. (2022). A review of human physiological responses to light: implications for the development of integrative lighting solutions. LEUKOS, 18(3), 387-414.

Published

24/06/2022

How to Cite

PAULA, P. L. A. H. de; PRAZERES, C. L. S. dos; CAMPOS, L.; FERREIRA, C. C. Evaluation of the illuminance on the monitor screen present in the hemodynamics room. Research, Society and Development, [S. l.], v. 11, n. 8, p. e39711831172, 2022. DOI: 10.33448/rsd-v11i8.31172. Disponível em: https://rsdjournal.org/index.php/rsd/article/view/31172. Acesso em: 27 nov. 2024.