How we process flexible endoscopes - Experience report

Authors

DOI:

https://doi.org/10.33448/rsd-v10i8.17126

Keywords:

Disinfection; Endoscopy; Hospital nursing service; Nursing; Equipment contamination.

Abstract

Objective: to report on how we carry out the cleaning process of flexible endoscopes. Method: a descriptive study, experience report type, about the cleaning and disinfection process of flexible endoscopes in a teaching hospital in southern Brazil. Results: after using the flexible endoscope, start the leak test in the examination room right after the end of the procedure, together with a pre-cleaning of the device in a vat of diluted enzymatic detergent solution. Visual inspection should be performed prior to performing high-level disinfection, according to its classification as a critical accessory, used in invasive procedures with penetration of the skin, mucous membranes, sterile spaces or cavities, sub-epithelial tissues and vascular system. After manual cleaning, the automated cleaning process is performed, including rinsing and high-level disinfection. Then the device is identified on the processing carried out and stored in cabinets in the vertical position. Conclusion: the reprocessing of flexible endoscopes must follow recommendations based on evidence in updated guidelines, investment in the physical structure with correct flows and storage of equipment in a manner appropriate to the reality of the service to reduce the rate of infection.

Author Biographies

Onilda Rubin, Hospital de Clinicas de Porto Alegre

Enfermeira.

Bruna Noschang Brum, Universidade Federal do Rio Grande do Sul

Enfermeira. Mestranda em Enfermagem pela Universidade Federal do Rio Grande do Sul

Lisiane Paula Sordi Matzenbacher, Hospital de Clinicas de Porto Alegre

Enfermeira. Mestranda em Saúde Coletiva pela Universidade Federal do Rio Grande do Sul. Pós graduação em Centro Cirúrgico e Sala de Recuperação Anestésica, especialista em Gestão em Saúde, Unidade de Terapia Intensa Adulto .

Ana Karina Silva da Rocha Tanaka, Universidade Federal do Rio Grande do Sul

Enfermeira. Doutora. Professora Adjunta da Escola de Enfermagem da Universidade Federal do Rio Grande do Sul

Ana Maria Pagliarini, Universidade Federal do Rio Grande do Sul

Acadêmica de enfermagem da Universidade Federal do Rio Grande do Sul

Débora Machado Nascimento do Espírito Santo, Hospital de Clinicas de Porto Alegre

Enfermeira

Carina Galvan, Hospital de Clinicas de Porto Alegre

Enfermeira.Especialista em Centro cirúrgico e sala de recuperação anestésica e em ortopedia e traumatologia.
Pós graduação em pediatria e neonatologia

References

Barbosa, M. C. T., Pinto, A. C. S., Alcantara, L. F. F. L., Macedo, E. C., Teixeira, V. M. F., Couto, L. L., et al. (2021). Helth Educacion through social media for patients undergoing digestive endoscopy: Integrative review. Research, Society and Development, 10(3), 4710312983. 10.33448/rsd-v10i3.12983.

Brasil. ANVISA. (2012). Resolução - RDC nº 15, de 15 de março de 2012. Dispõe sobre requisitos de boas práticas para o processamento de produtos para saúde e dá outras providências. 2012. Disponível em: https://bvsms.saude.gov.br/bvs/saudelegis/anvisa/2012/rdc0015_15_03_2012.html.

Brasil. ANVISA (2013). Resolução- RDC nº 6, de 10 de março de 2013. Dispõe sobre os requisitos de Boas Práticas de Funcionamento para os serviços de endoscopia com via de acesso ao organismo por orifícios exclusivamente naturais. https://bvsms.saude.gov.br/bvs/saudelegis/anvisa/2013 /rdc0006_10_03_ 2013.html11.

Feitosa, K.A., Ferraz, C. R. & Feitosa, F. P. A. (2020). Compreensão da Equipe de Enfermagem de Uma Central de Material Frente ao Riscos Ocupacionais Relacionado ao Processo de Limpeza dos Artigos Médicos. Revista JRG de Estudos Acadêmicos, 3(7).

Gomes B. P., M., Guadagnin, S. V. T., Azevedo, A. da S., Costa L., C. da, Costa, D. de M., & Ferreira Veiga Tipple, A. (2020). Experiência na implantação de um protocolo para processamento de endoscópios adotando uma metodologia ativa. Comunicação Em Ciências Da Saúde, 31(01), 189-196. 10.51723/ccs.v31i01.622.

Graziano, K. U., Pereira, M. E. A. & Koda, E. (2016). Proposta metodológica para a validação da eficácia de desinfecção de processadora automática de endoscópios flexíveis. Rev. Latino-Am. Enfermagem. 10.1590/1518-8345.0595.2745.

Kenters, N., Huijskens, E. G. W., Meier, C. & Voss A.(2015). Infectious diseases linked to cross-contamination of flexible endoscopes. Endosc Int Open, 3:259-65. 10.1055/s-0034-1392099.

Kovaleva, J. Infectious complications in gastrointestinal endoscopy and their prevention. (2016). Best Pract Res Clin Gastroenterol., 30(5):689-704. 10.1016/j.bpg.2016.09.008.

Mati, M. L. M., Guimarães, N. R., Magalhães, P. P., Farias, L. M. & Oliveira, A. C.(2019). Enzymatic detergent reuse in gastroscope processing: a potential source of microorganism transmission. Revista Latino-Americana de Enfermagem, 27, e3211. 10.1590/1518-8345.3101.3211.

Nascimento, A. C., Cunha, C. I. J., Silva, C. R. G., Leão, M. V. P., Santos, S. S. F. (2015). Estabilidade do ácido peracético no processo de desinfecção prévia à lavagem. Rev. Assoc. Paul. Cir. Dent, 69(.4). 367-382.

Oliveira, A.C. & Mati, M. L. M. (2015). Fatores relacionados à troca das soluções de desinfecção dos aparelhos endoscópicos Rev. SOBECC, 20(1): 24-9. 10.5327/Z1414-4425201500010006.

Oliveira, C. L., Silva, D. M. & Bopsin, P. S.(2018). Desinfecção de endoscópios: revisão integrativa. Anais da XI Mostra Científica do CESUCA, 11:390-408. Disponível em: http://ojs.cesuca.edu.br/index.php/mostrac/article/view/1390.

Paczek, R. S., Espírito Santo, D. M. N. & Galvan, C. (2020). Utilização de equipamentos de proteção individual em unidade endoscópica.Rev enferm UFPE on line. 2020;14:e243993. 10.5205/1981-8963.2020.243993.

Primo, M. G. B., Guadagnin, S. V. T., Azevedo, A. S., Luciano, C. C., Costa, D. M., & Tipple, A. F. V. (2020). Experiência na implantação de um protocolo para processamento de endoscópios adotando uma metodologia ativa. Comunicação Em Ciências Da Saúde, 31(01), 189-196. 10.51723/ccs.v31i01.622.

Rahman, M. R., Perisetti, A., Coman, R. M., Bansal, P., Chhabra, R. & Goyalet, H. (2019). Duodenoscope-Associated Infections: Update on an Emerging Problem. Dig Dis Sci, 64: 1409–18. 10.1007/s10620-018-5431-7.

Ribeiro, M. M., Oliveira, A. C., Ribeiro, S. M., Watanabe, E., Stoianoff, M. A. R. & Ferreira, J. A.(2013). Eficácia do reprocessamento de endoscópio gastrointestinal flexível. Infect Control Hosp Epidemiol, 34 (3): 309 - 12. 10.1086/669518.

Sobecc (2017). Diretrizes de Práticas em Enfermagem Cirúrgica e Processamento de Produtos para a Saúde. Rev. SOBECC. (7a ed.), Manole:135-46.

Schwab, K. & Singh, S. (2011). An introduction to flexible endoscopy. Surgery, 29 (2): 80–4. 10.1016/j.mpsur.2010.11.014.

Silva, J. M. B., Loureiro, L. H., Silva, I. C. M. & Novaes, M. L. (2020). Coronavirus and the disinfection and reprocessing protocols of hospital articles. Research, Society and Development, 9(9), e29996187. 10.33448/rsd-v9i9.6187.

WGO. World Gastroenterology Organization (2019). Endoscope disinfection update: a guide to resource-sensitive reprocessing. USA: World Gastroenterology Organization. Disponível em: https://www.worldgastroenterology.org/UserFiles/file/guidelines/endoscope-disinfection-english-2019.pdf.

Published

05/07/2021

How to Cite

RUBIN, O. .; PACZEK, R. S. .; BRUM, B. N.; MATZENBACHER, L. P. S. .; TANAKA, A. K. S. da R.; PAGLIARINI, A. M.; SANTO, D. M. N. do E. .; GALVAN, C. How we process flexible endoscopes - Experience report. Research, Society and Development, [S. l.], v. 10, n. 8, p. e6810817126, 2021. DOI: 10.33448/rsd-v10i8.17126. Disponível em: https://rsdjournal.org/index.php/rsd/article/view/17126. Acesso em: 24 dec. 2024.

Issue

Section

Health Sciences